MANUAL # OMP L80-P
REV. 4
INTRODUCTION
The K-Dyne, Inc. pneumatic
version of the Model L80 is a 2" NPT mounted liquid level switch. It
is typically mounted in a K-Dyne Model C80 cage shown in the Figure
1 for external bridle mounting or may be internally mounted in a 2"
NPT (F) connection. Both mounting configurations are shown in Figure
2.
These sensors
send or remove a pneumatic signal when a liquid in a tank or a
pressurized vessel reaches a predetermined level. The L80 operation
mode as either a low (direct) or high (reverse) acting level switch
is selected by rotating the pivot body 180˚. The switch functions as
a three way 'BLOCK and BLEED' valve. When used as a low acting
switch the valve will bleed off a pneumatic signal when the liquid
level is below the sensor trip point. For a high acting switch, the
valve bleeds off a signal when the liquid level is above the sensor
'trip' point.
2.0 PRINCIPLE OF
OPERATION (Refer to Figure 3)
The pneumatic version of
the K-Dyne, Inc. Model L80 liquid level switch is a multi-functional
pneumatic liquid level sensor. Depending on the mode of operation,
it functions as a three way, normal closed (N.C.) or normally open
(N..O.) pneumatic valve that will operate when the liquid level
rises or falls below a fixed point in a tank or pressurize vessel.
Valve supply pressures may be from 10 p.s.i. to 35 p.s.i. and 30
p.s.i. is recommended for the most efficient operation.
The switch senses a change
in the buoyancy of a displacer assembly (9) inside either a
pressurized or an unpressurized vessel. A pivoting action is used to
transmit the buoyancy change of the displacer assembly. The bore of
the cone (14) contacts a pivot (13) on the pivot body (3). The
o-ring (8) provides a pressure tight seal between the body (7) and
the shaft of the cone protruding through the o-ring.
Pressure in the vessel
acting on the shaft of the cone forces the cone against the pivot.
The position of the pivot point, at the center of the o-ring, allows
the displacer assembly to move vertically with the liquid level.
This motion is guided to act along the axis of the transverse rod
(15) by two vertical tabs on the pivot body that mate with notches
in the cone. Any vertical motion of the displacer is transmitted by
the cone to the ends of the transverse rod. A ramp on the outside
diameter of the transverse rod contacts the tip of the valve stem
(2G) to operate the valve.
2.1 LOW
OPERATION
For use as a low-level
sensor (output on rising level), the pivot body (3) is orientated
with the enlarged end of the transverse rod (15) in the up position,
towards the spring cap (4). The displacer assembly weight exerts a
clockwise rotational force on the cone through the pivot. The spring
(5), acting through he spring guide (6) exerts a counterclockwise
rotational force on the cone. When the liquid level is below the
displacer assembly, the clockwise force is greater. This greater
force causes the cone to push upward against the transverse rod and
the seat area of the valve stem (2G) remains positioned away from
the ball (2C). In this position the ball is in contact with the seat
(2L) and pressure at the 'SUPPLY' port is prevented from entering
the 'OUTPUT' port. The 'OUTPUT' port and 'VENT' port are in
communication allowing pressure at the 'OUTPUT' port to be
vented.
As the liquid level rises
and begins to cover the displacer assembly, its buoyancy, in the
liquid, reduces the clockwise rotational force on the cone. When the
spring exerts a greater counterclockwise rotational force the cone
moves the transverse rod downward and the transverse rod ramp pushes
the valve stem toward the ball. As the valve stem contacts the ball,
the 'VENT' port is blocked and the ball disengages the seat.
Pressure form the 'SUPPLY' port then passes thought the valve body
to the 'OUTPUT' port. 2.2 HIGH OPERATION
For use as a high level
sensor (output on falling level), the pivot body is oriented with
the enlarged end of the transverse rod (15) away from the spring cap
(4). When the liquid level is above the displacer assembly, the
counterclockwise rotational force of the spring on the cone is
greater than the counterclockwise rotational force of the buoyant
displacer. This larger force maintains the transverse rod in the
down position causing the transverse rod ramp to keep the ball (2C)
against the valve stem (2G) and away from the seat (2L).
Communication between the 'VENT' port and 'OUTPUT' port is blocked
and pressure from the 'SUPPLY' port is allowed to pass to the
'OUTPUT' port.
Figure 3
As the liquid level falls,
the displacer loses its buoyancy and the clockwise rotational force
on the cone becomes greater than the counterclockwise rotational
force of the spring. The transverse rod is shifted upward allowing
the valve stem to follow the transverse rod ramp, positioning the
ball against the seat and blocking pressure communication between
the 'SUPPY' and 'OUTPUT' ports. As the ball contacts this seat the
tip of the valve stem loses contact with the ball and allows
pressure at the 'OUTPUT' port to pass thought the valve 'VENT'
port. 3.0
INSTALLATION
3.1 EXTERNAL CAGE TO VESSEL
It is suggested that the
Model L80 be installed in and external cage for maintenance and
testing purposes. A recommended piping system for mounting an
external cage to a vessel is shown in figure 2. The valves above and
below the cage are required for testing and maintenance
purposes.
If a cage with butt weld
connections is used, the level switch should not be installed in the
cage before welding. This will prevent weld sparks form damaging the
displacer assembly. After welding install the level switch as per
instructions in Section 3.2, LEVEL SWITCH INSTALLATION.
If the cage has threaded or
flanged process connections, the level switch does not have to be
removed before mounting the switch to the vessel. Cages with 1" NPT
process connections may be installed on the vessel with 1", ¾", or
½" O.D. piping. CAUTION: The pressure rating of all pipe, valve, and
pipe fittings must meet or exceed the working pressure of the vessel
they are to be installed on. Adequate support for long pipe runs
should be provided to prevent excessive vibrations of the level
switch.
It is recommended that
external cages be installed with the process connections as close to
vertical and the cage length as close to horizontal as possible.
This will insure the proper fluid flow through the cage.
3.2 LEVEL SWITCH
INSTALLATION
Before installing a level
switch into an external cage, tank or vessel verify that the level
switch is configured for the operation mode required (i.e. direct
(low) or reverse (high) action). See Section 4.0 CHANGING MODE OF
OPERATION and figure 5. NOTE: When direct mounting a level switch to
a tank or vessel verify that the back of the displacer assembly will
be at least ½" (13mm) longer than the 2" connection it is being
installed into. (See figure 4)
TOOL REQUIRED: 2.125" open
end or suitable adjustable wrench or pipe wrench
| 3.2.1 |
Clean any dirt or debris form
the 2" NPT pipe threads of the switch and mating
connection |
| 3.2.2 |
Apply Teflon tape or other
thread sealing compound to the 2" NPT threads on the
switch |
| 3.2.3 |
Apply and anti-galling compound to the 2" NPT female
thread that the switch is to be installed into. |
| 3.2.4 |
Screw the switch into the mounting tread and tighten
with the wrench. When tight, the spring cap must be up as
shown in figure 5. |
CAUTION: If using
a pipe wrench do not allow the wrench jaws to contact the spring
cap. Wrench forces may damage the cap and prevent the switch from
operating properly.
NOTE: THE SPRING CAP MUST
ALWAYS BE IN THE VERTICAL POSITION AS SHOWN IN FIGURE 5 FOR THE
SWITHCH TO OPERATE PROPERLY.
| 3.2.5 |
Orientate the 'OUTPUT' port to
facilitate connection of instrument lines by loosening the
three set screws (2F) on the valve body (2A) and rotating the
valve body to the desired position.
|
CAUTION: DO NOT position
the valve body 'VENT' port in the up position. This allows debris
and liquids to accumulate in the valve assembly and prevent proper
operation. Align the 'VENT' port in either the horizontal or down
position.
| 3.2.6 |
Lightly tighten each of the
three set screws first and then firmly tighten them in a
clockwise pattern. |
4.0 CHANGING MODE OF
OPERATION (Refer to figure 5)
TOOL REQUIRED: 3/16" Allen
wrench
The level switch mode of
operation may be easily changed. If the level switch is installed in
a pressurized cage or vessel, the pressure does not have to be
removed. Figure 5 shows the proper orientation of the level switch
function stamping for low and high acting operation.
4.1 UPRESSURIZED CAGE OR
VESSEL
| 4.1.1 |
Verify that no pressure is
present in the cage or vessel. If this cannot be verified,
follow instructions in Section 4.2 PRESSURIZED CAGE OR
VESSEL. |
| 4.1.2 |
Block and bleed the instrument
pressure from the 'SUPPLY' port. Remove any instrument tubing
connected to the 'SUPPLY' and 'OUTPUT' ports. |
| 4.1.3 |
With the 3/16" Allen wrench remove the four hex socket
head cap screws. (1). CAUTION: If the pivot body moves outward
as these screws are being removed, the vessel or cage is not
completely depressurized. Re-install the screws and perform
procedures in Section 4.2 of this manual. Rotate the pivot
body 180˚ to change the mod of operation from low to high or
form high to low as desired. See figure 5 for proper pivot
body orientation. |
| 4.1.4 |
Replace the four hex socket head cap screws and
tighten. |
| 4.1.5 |
Re-align the valve ports as desired using the
procedures given in paragraph
3.2.5. |
4.2 PRESSURIZED CAGE OR
VESSEL
| 4.2.1 |
Block and bleed the instrument
pressure form the 'SUPPLY' port. Remove any instrument tubing
connected to the 'SUPPLY' and 'OUTPUT' ports |
| 4.2.2 |
Use a 3/16" Allen wrench and
remove two of the hex socket head cap screws (1) that are
diagonally apart. Replace the two hex socket head cap screws
just removed with two 1/4" -20 screws or bolts that are 1-1/4"
to 1-1/2" long. These serve as a safety stop in the event the
level switch has received unknown damage. Rotate the remaining
two screws on turn at a time, alternating between the two. The
process pressure, acting on the end of the cone (14) that is
protruding through the body, will move the pivot body outward
as the two screws are turned. When the pivot body has stopped
moving, remove the four screws completely. Rotate the pivot
body 180˚ to change the mode of operation if necessary. See
figure 5 for proper pivot body
orientation. |
CAUTION: THE PIVOT BODY
SHOULD MOVE OUTWARD APPROXIMATELY 1/16" AND STOP AS THE ROD (11)
CONTACTS THE BODY. IF THE PROCESS PRESSURE IS STILL EXERTING A FORCE
ON THE PIVOT AFTER IT MOVES 1/8" STOP TURNING THE TWO SCREWS. THE
UNIT MAY HAVE DAMAGE THAT WILL ALLOW THE CONE TO DISENGAGE THE
O-RING (8) AND THE PROCESS PRESSURE AND MEDIUM WILL ESCAPE THROUGH
THE BODY. THE PROCESS PRESSURE WILL HAVE TO BE REMOVED FROM THE CAGE
OR VESSEL BEFORE ANY FURTHER DISASSEMBLY IS POSSIBLE.
| 4.2.3 |
Replace the four hex
socket head cap screws and tighten. |
 |
| 4.2.4 |
Reconnect the instrument
lines. |
5.0 DISSASSEMBLY (Refer to
figures 1 & 3)
TOOLS REQUIRED:
- 7/16" open end wrench or
adjustable wrench
- 5/18" open end wrench or
adjustable wrench
- 3/16" and 1/8" Allen
wrenches
- needle nose
pliers
- pliers
- o-ring pick or small
screwdriver
CAUTION: ONLY THE VALVE
ASSEMBLY AND PIVOT BODY CAN BE REMOVED FROM THE LEVEL SWITCH UNIT
WITHOUT DEPRESSURIZING THE CAGE OR VESSEL. ANY FURTHER DISASSEMBLY
WITH PRESSURE PRESENT ON THE CAGE OR VESSEL WILL RESULT IN SEVERE
PERSONAL INJURY OR DAMAGE TO EQUIPMENT.
5.1 VALVE
DISASSEMBLY
| 5.1.1 |
Bleed off all instrument
pressure from the valve assembly and remove all tubing and
fittings from the valve body (2A). |
| 5.1.2 |
Remove the valve assembly from
the pivot body (3) by rotating it counterclockwise. NOTE: The
valve stem (2G) and stem spring (2H) are free to move and care
should be taken not to accidentally let them separate from the
valve assembly. To avoid damaging or marring the valve
assembly it is recommended to install a pipe fitting into the
'VENT' poet of the valve assembly and tap it to remove the
valve. |
| 5.1.3 |
Pull the valve stem and stem spring out of the adapter
head (2J). |
| 5.1.4 |
To remove the filter disc (2M) from the valve body,
insert the tip of an O-ring pick or small screwdriver between
the outside diameter of the filter disc and the inside
diameter of the 'SUPPLY' port. Pry inward and upward with the
pick or screwdriver to rotate the disc edgewise in the port.
Use needle nose pliers to pull the disc from the hole.
|
| 5.1.5 |
Loosen the three set screws (2F) sufficiently enough to
allow the adapter head to be pulled from the valve body (2A).
It is not necessary to completely remove these
screws. |
| 5.1.6 |
Insert an Allen wrench or other rod shaped tool thru
the 'SUPPLY' port and push the ball (2C) and valve seat (2L)
out of the valve body. |
| 5.1.7 |
Use a pick or small screwdriver to remove the o-ring. This is permanently installed in the valve body. O-rings
(2D) & (2E) from the body and O-rings (2D) & (2K) from
the seat. Discard all used seals. |
| 5.1.8 |
DO NOT remove the ball retainer (2B) from the valve body. |
| 5.1.9 |
The valve stem wear ring should not be removed unless
damaged or an excessive amount of debris has accumulated on
the O.D. o f the valve stem. These rings are split and may be
removed like and O-ring. The wear rings are plastic and
breakable, exercise care when removing them. |
| 5.1.10 |
The plastic cap on the valve stem should also not be
removed and attempting to do so would change the factory set
adjustment. Switch re-calibration would be
necessary. |
5.2 PIVOT BODY
DISSASSEMBLY
| 5.2.1 |
For pivot body removal from a
non-pressurized cage or vessel use a 3/16" Allen wrench and
remove the four hex socket head cap screws (1). |
| 5.2.2 |
For pivot body removal from a
pressurized cage or vessel use a 3/16" Allen wrench and remove
the two of the hex socket head cap screws that are diagonally
apart. . Replace the two hex socket head cap screws just
removed with two 1/4" -20 screws or bolts that are 1-1/4" to
1-1/2" long. These serve as a safety stop in the event the
level switch has received unknown damage. Rotate the remaining
two screws one turn at a time, alternating between the two. The
process pressure, acting on the end of the cone (14) that is
protruding through the body, will move the pivot body outward
as the two screws are turned. When the pivot body has stopped
moving, remove the four screws
completely. |
CAUTION: THE PIVOT BODY
SHOULD MOVE OUTWARD APPROXIMATELY 1/16" AND STOP AS THE ROD (11)
CONTACTS THE BODY. IF THE PROCESS PRESSURE IS STILL EXERTING A FORCE
ON THE PIVOT AFTER IT MOVES 1/8" STOP TURNING THE TWO SCREWS. THE
UNIT MAY HAVE DAMAGE THAT WILL ALLOW THE CONE TO DISENGAGE THE
O-RING (8) AND THE PROCESS PRESSURE AND MEDIUM WILL ESCAPE THROUGH
THE BODY. THE PROCESS PRESSURE WILL HAVE TO BE REMOVED FROM THE CAGE
OR VESSEL BEFORE ANY FURTHER DISASSEMBLY IS POSSIBLE.
| 5.2.3 |
Pull the pivot body (3) from
the body (7). NOTE: The transverse rod (15) fits loosely in
the pivot body and care should be taken not to accidentally
let if fall out of the pivot body. |
| 5.2.4 |
Remove the transverse rod form
the pivot body |
| 5.2.5 |
Use pliers to hold the O.D. of
the pivot point (13) and rotate it counter clockwise to remove
it form the pivot body. |
5.3 FULL
DISSASSEMBLY
| 5.3.1 |
For full disassembly, it is
necessary to completely remove the pressure form the cage or
vessel the switch is installed in. All liquids should be
drained to a position below the level
switch. |
CAUTION: ALL PRESSURE MUST
BE REMOVED FROM THE CAGE OR VESSEL BEFORE FULL DISASSEMBLY CAN BE
PERFORMED. FAILURE TO REMOVE THE PRESSURE MAY RESULT IN PERSONAL
INJURY OR DAMAGE TO EQUIPMENT.
| 5.3.2 |
Loosen and remove the level
switch from the tank, vessel or cage that it is mounted
to. |
| 5.3.3 |
Hold the rod (11) with the pliers and rotate it
counter-clockwise to remove it from the level switch.
|
| 5.3.4 |
Normally it is not necessary
to remove the displacer assembly from the rod. If required,
loosen the lock nut (10) and unthread the displacer assembly
from the rod. |
| 5.3.5 |
Follow instructions in Section 5.1 VALVE DISASSEMBLY to
remove and disassemble the valve body. |
| 5.3.6 |
Use the 5/8" wrench and rotate the spring cap (6)
counter clockwise to remove it from the body (5). Care should
be taken no to lose the spring (7) and spring guide
(8). |
| 5.3.7 |
Push the threaded end of the
cone (14) through the body. |
| 5.3.8 |
Using an o-ring pick or small screwdriver remove the
O-ring (8) form the body. |
6.0 REPAIR AN ASSEMBLY
REFER TO THE SWITCH DATA SHEET FOR REPLACEMENT PART
NUMBERS
TOOLS REQUIRED:
- 7/16" open end wrench or
adjustable wrench
- 5/18" open end wrench or
adjustable wrench
- 3/16" and 1/8" Allen
wrenches
- needle nose
pliers
- pliers
- safety solvent
- silicone based o-ring
lubricant
| 6.1 |
Using an appropriate safety
solvent, clean all parts. |
| 6.2 |
Inspect the I.D. of the
transverse rod bore in the pivot body (3) and the O.D. of the
transverse rod (15) for corrosion and excessive wear. The
surface of the bore and the ramp on the transverse rod should
have a mirror like appearance. Replace any worn or damaged
parts. |
| 6.3 |
Examine the valve stem (2G) for corrosion and excessive
wear. The small sealing diameter should have a mirror like
surface and the seat at the tip should be free of
indentations. Insure that the cap that contacts the transverse
rod ramp is still pointed. Replace any worn or damaged
parts. |
| 6.4 |
The seat (2L) at the contact point of the ball (2C)
should be free of any indentations. The ball should also have
a polished finish and have no corrosion or gouges.
|
| 6.5 |
If the level switch has a stainless steel displacer
assembly, it should be free of dents, cracks or
holes. |
| 6.6 |
Lubricate all seals from the repair kit with a silicone
based lubricant. |
| 6.7 |
Lightly lubricate the filter disc (2M) O-ring and push
the disc edgewise into the center of the 'SUPPLY' port until
it stops. Start rotation the disc across the port so the
filter screen will be toward the ball retainer (2B) while
pushing downward on the outside edge that is closest to the
end of the valve body. |
| 6.8 |
Use a blunt rod, such as an Allen wrench, and push the
edge of the disc as far down a possible. Using the end of an
O-ring pick or small screwdriver inserted between the disc
O.D. and 'SUPPLY' port I.D., press firmly on the O-ring around
the circumference of the filter disc to completely seat it in
the port. |
| 6.9 |
Install the O-rings (2D) and (2E) into the valve body
and the O-rings (2D) and (2K) on the seat. O-ring (2K) is a
blue color. |
| 6.10 |
Place the ball into the valve body and install the
seat. Place the adapter head (2J) into the valve body and
tighten he set screws (2F). |
| 6.11 |
Lubricate the valve stem and transverse rod. DO NOT
excessively lubricate, this will prevent proper operation of
the level switch. Be sure to re-install the wear rings if
removed. |
| 6.12 |
Install the valve stem and stem spring (2H) into the
adapter head (2J). Depress the valve stem down into the body
and release. It should move completely and freely outward.
NOTE: Care should be taken not to lose these pieces while
handling. |
| 6.13 |
Insert the small diameter of the transverse rod (15)
into the counter bored side of the hole of the pivot body (3),
and push it full into the pivot body. Slide it in and out
approximately 1/8" to verify that it is not
binding. |
| 6.14 |
Thread the valve into the pivot body. To tighten the
valve assembly and to avoid damaging or marring the valve
assembly it is recommended to install a pipe fitting into the
'VENT' port of the valve assembly and tap it
lightly. |
| 6.15 |
Install the body O-ring (8) into the body
(7). |
| 6.16 |
Thread the replacement pivot (13) onto the pivot body
and tighten. Lubricate the pivot point and the square tabs on
the pivot body. |
| 6.17 |
Lubricate the shaft and the treads on the treaded end
of the cone. Install the cone into the body by turning the
cone clockwise and pushing gently. |
| 6.18 |
Align the tabs on the pivot body with the slots in the
cone (14) and slide the pivot body into the level switch
body. |
| 6.19 |
Align the valve body in the correct position for the
desired direct or reverse operation, see figure 5. Install and
tighten the four hex socket and head cap screws
(1). |
| 6.20 |
Thread the rod (11) onto the cone and tighten securely
with pliers. If the displacer assembly was removed, reinstall
and tighten the lock nut (10) securely. |
| 6.21 |
Insert the spring (5) and spring guide (6) into the
spring cap (4). Install the spring cap into the body and
tighten. |
| 6.22 |
Replace the level switch assembly into the cage or
mating flange as per instructions in Section 3.0
INSTALLATION. |
7.0 SWITCH
CALIBRATION
| 7.1 |
Due to manufacturing tolerances or part wear, it may be
necessary to adjust the valve stem (2G) length when a new or
repaired valve assembly (2) is installed. |
| 7.2 |
Install the repaired or new
valve assembly per procedures in paragraph 6.14 of this
manual. |
| 7.3 |
Connect a suitable instrument supply to the valve
assembly 'SUPPLY' port and an indication device such as a
pressure gauge to the 'OUTLET' port. |
| 7.4 |
If the switch is not installed in a vessel it may be
tested by raising and lowering the displacer assembly
manually. For a switch installed in a vessel or external cage
it is recommended to use the actual process fluid to check for
proper switch operation. |
| 7.5 |
Switches that are installed
but not practical to operate with the process fluids may be
tested by removing the spring cap (4) and the spring (5) then
depressing and releasing the spring guide (6). |
| 7.6 |
Should the switch operation have one of the malfunction
s as given in the table of 7.7, use procedures given in
paragraphs 7.8 thru 7.11 to calibrate the switch. |
| 7.7 |
| Switch
Response |
Calibration |
| Output signal
does not bleed off |
Shorten Valve Stern |
| Slow bleed off
of output signal |
| No output
signal |
Lengthen Valve Stern |
| Slow increases
of output signal | |
| 7.8 |
Remove the valve assembly, valve stem and stem spring
per procedures in paragraphs 5.1.2 and 5.1.3. |
| 7.9 |
Insert a small dowel, pick or Allen Wrench into the
hole on the plastic cap of the valve stem and one in the cross
drilled hole of the stem. With the plastic cap facing up,
rotate the cap approximately 30˚ counterclockwise to lengthen
and clockwise to shorten the stem. |
| 7.10 |
Re-install the valve assembly onto the switch and then
retest the switch. |
| 7.11 |
Repeat these procedures as required until the switch is
functioning properly. |
8.0 RECOMMENDED
MAINTENANCE
| PROCEDURE |
INTERVAL |
| Test switch in place with liquid to check
for proper operation. |
every 30
days |
| Clean vent ports of debris |
every 30
days |
| Replace body seal (8) |
as
required |
TROUBLE
SHOOTING:
| PROBLEM |
POSSIBLE CAUSES |
RECOMMENDED ACTION |
| Level Switch does not
function |
Debris blocking the valve vent
port |
Clean vent port and test
switch. |
|
Debris on the inside of the
cone |
Remove
pivot body (3) per procedures in Section 4.0 and clean cone
I.D. |
|
Valve not functioning |
Follow
procedures in section 5.2 and 6.0 to disassemble and clean
valve. Replace seals if necessary. |
|
Supply port plugged |
Disconnect all instrument tubing and clean supply port.
Clean instrumentation system filters. |
|
Connection tubing plugged |
Remove
and clean or replace instrument tubing. Clean instrumentation
system filters. |
|
Debris accumulation on displacer assembly
(9) or in external cage. |
Remove
switch from vessel or cage per procedures in Section 6.0 and
clean. |
|
Level switch not installed properly.
Spring cap not in the 'UP' position |
Use
instructions given in Section 3.0 & 4.0 to properly
install the switch. |
|
Specific gravity of fluid too
light for switch to function |
Consult factory for optional Models
available |
|
Stainless steel displacer assembly (if
used) damaged |
Remove
switch from vessel or cage per procedures in 5.3 and replace
displacer assembly. |
| Constant gas leakage from vent port in
valve |
Valve seals (2K) & (2D) damaged or
worn |
Follow
procedures in Sections 5.1 & 6.0 to disassemble and
replace the seals. |
|
Debris on valve seat (2L) or valve stern
(2G) |
Follow
procedures in Sections 5.1 & 6.0 to disassemble and clean
valve. |
| Liquid or gas leakage from
body vent |
Body o-ring (8) damaged
or worn |
Replace body seal (8) per instructions in
Sections 5.0 & 6.0. |
|